Connecting Calibration Gas Tanks & Mixing Chambers
When using calibration gas tanks with a mixing chamberāsuch as the AFT15 or the GASSYS2 systemāan interface is required between the gas tank barbed ports and the mixing chamberās Luer lock female port. Here is a suggested method.
- Acquire 3-6 feet of 1/8″ ID silicon tubing ANDĀ a 1/8ā barb to male Luer connector (polycarbonate).

- Assemble the male Luer-lock into one end of the silicon tubing.
- Push the other open end of the tubing into the barb port of the gas regulator.
- Interface theĀ AFT20 filter disk end and without the Y-connector that comes with it, to the male Luer-lock connector.
- Connect the AFT20 male Luer-lock to the female Luer-lock port of the mixing chamber.