Description
Impedance calibrator with 20 Ξ© reference and 50 Ξ© reference for the EBI100C, NICO100C, or NICO100C-MRI impedance amplifier.
ZCAL20 (20-ohm reference) and ZCAL50 (50 0hm reference) impedance calibrators plug into the EBI100C, NICO100C, or NICO100C-MRI via Touchproof pins.
- ZCALC is not compatible with the BioNomadix BN-NICO impedance amplifier or the NICO100D Smart Amplifier.
Calibration Procedure:
- Plug ZCAL20 into the front panel of the EBI100C, NICO100C, or NICO100C-MRI.
- Switch the EBI100C/NICO100C/NICO100C-MRI to the desired mode and confirm that the βZβ channel HP filter βDCβ option is selected.
- Record data on βZβ or impedance channel.
- Press CAL1 in AcqKnowledge and thenΒ map the resulting voltage to β20 ohms.β
- Remove ZCAL20 and replace with ZCAL50.
- Press CAL2 in AcqKnowledge and then map the resulting voltage to β50 ohms.β